Technical Specifications
Infusion rate range
Adjustable drop rate range: 1 /min~400 drops/min
(step: 1 drop /min)
Adjustable flow rate range: 1ml/h~1200ml/h
(step 0.1ml/h)
Purge rate: 800ml/h
KVO rate: 1ml/h (less than 100ml/h)
2ml/h (100~300ml/h)
3ml/h (greater than 300ml/h)
Infusion volume range: 1ml~9999.9ml (step: 0.1 ml)
Maximum accumulated volume: 9999.9ml (step: 0.1 ml)
Infusion Time range: 1min~9999min (step: 1 min)
Infusion accuracy: Within ±3% (using calibrated high quality IV set)
Mechanical precision: within ±2%
Occlusion alarm threshold:
High: 800 mmHg ± 200 mmHg (106.7kPa±26.7kPa)
Medium: 500 mmHg ± 100 mmHg (66.7kPa±13.3kPa)
Low: 300 mmHg ± 100 mmHg (40.7kPa±13.3kPa)
Alarm: Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, Idle.
Size: 188 mm(L)×198 mm(W)×228 mm(H)
Net weight: 2.5kg